Adam Equipment WBW Series Short Manual
Adam Equipment WBW Series Short Manual

Adam Equipment WBW Series Short Manual

Hide thumbs Also See for WBW Series:

Advertisement

Adam Equipment
WBW & WBW-M SERIES
(P.N. 3036610540, Revision A1, Oct 2009)
Adam Equipment strives to be more environmentally focused and uses recycled materials and
environmentally friendly packaging where possible. As part of this initiative we have developed a short
manual that uses less paper and ink to describe the main functions of your new Adam indicator/scale. A
complete version is available at www.adamequipment.com. Thank you for your support of Adam Equipment
and we hope that you enjoy your new scale.
© Adam Equipment Company 2009

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the WBW Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Adam Equipment WBW Series

  • Page 1 Adam indicator/scale. A complete version is available at www.adamequipment.com. Thank you for your support of Adam Equipment and we hope that you enjoy your new scale.
  • Page 2 © Adam Equipment Company 2009...
  • Page 3: Table Of Contents

    SETTING UP AND TURNING ON THE SCALES ........2 OPERATION ..................2 NUMERIC ENTRY METHOD .............2 ZEROING...................2 TARING .....................2 WEIGHING..................3 CHECK-WEIGHING................3 RS-232 SPECIFICATION..............4 CALIBRATION..................4 ERROR CODES..................5 SPECIFICATIONS ................6 ERROR MESSAGES ................9 MENU STRUCTURE................10 WARRANTY INFORMATION .............11 10.0 SERVICE INFORMATION ...............12 P a g e © Adam Equipment Company 2009...
  • Page 4: Setting Up And Turning On The Scales

    WBWa series are usually set for imperial units. SETTING UP The WBW Series comes with a stainless steel pan. Place it on the top if already not installed. Do not press with excessive force as this could damage the load cell inside.
  • Page 5: Weighing

    If the scale is turned off it stores the last high and low limits in the memory along with information about the weighing unit in use when the limits were stored. If the scale is turned on again, the limits and the weighing unit will be active. P a g e © Adam Equipment Company 2009...
  • Page 6: Rs-232 Specification

    Press the [Tare] key when the stable indicator is on. The scale will calibrate to the mass and then return to weighing. Remove the calibration weight as soon as calibration is complete. P a g e © Adam Equipment Company 2009...
  • Page 7: Error Codes

    Improper calibration (should be within +10% of the factory calibration). The old calibration data will be FAIL L FAIL L FAIL L FAIL L retained until the calibration process is complete. P a g e © Adam Equipment Company 2009...
  • Page 8: Specifications

    4000 g 8000 g 16000 g Tare Range -2000 g -4000 g -8000 g -16000 g Readability 0.2 g 0.5 g Repeatability (S.D.) 0.2 g 0.5 g 0.4 g Linearity (±) P a g e © Adam Equipment Company 2009...
  • Page 9 Stabilisation Time 2 seconds typical Operating Temperature -10° C to 40° C Power supply 12 VDC, 800 Ma through an external Power Supply Module Battery Internal rechargeable battery (~50 hours operation) P a g e © Adam Equipment Company 2009...
  • Page 10 IP 65 Sealed ABS Plastic housing with Stainless Steel pan Pan Size 210 x 173 mm Overall Dimensions 231 x 265 x 153 mm (wxdxh) Net Weight 3.3 kg / 7.26 lb Applications Weighing Scales Functions Weighing, Check weighing P a g e © Adam Equipment Company 2009...
  • Page 11: Error Messages

    FAIL H or Calibration error Improper calibration (should be within +10% of the factory calibration). The old calibration data FAIL L will be retained until the calibration process is complete. P a g e © Adam Equipment Company 2009...
  • Page 12: Menu Structure

    MENU STRUCTURE P a g e © Adam Equipment Company 2009...
  • Page 13: Warranty Information

    WARRANTY INFORMATION Adam Equipment offers Limited Warranty (Parts and Labour) for components that fail due to defects in materials or workmanship. Warranty starts from the date of delivery. During the warranty period, should any repairs be necessary, the purchaser must inform its supplier or Adam Equipment Company.
  • Page 14: 10.0 Service Information

    Include any recent history of the unit. For example: -Has it been working since it’s delivered -Has it been in contact with water -Damaged from a fire -Electrical Storms in the area -Dropped on the floor, etc. P a g e © Adam Equipment Company 2009...
  • Page 15 Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device. Changes or modifications not expressly approved by Adam Equipment could void the user's authority to operate the equipment. WEEE COMPLIANCE...
  • Page 16 For a complete listing of all Adam products visit our website at www.adamequipment.com © Copyright by Adam Equipment Co. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment.

This manual is also suitable for:

Wbw-m seriesWbwa seriesWbw 3mWbw 6mWbw 2/5aWbw 15m ... Show all

Table of Contents