2. IF YOU REPLACE THE CIS UNIT
<What to do after replacement>
• Acquire white level data and set CIS scan area
• Checking operation after repair
What you need to prepare
(1) Calibration sheet (LD1914001)
(2) USB flash memory
(3) Scan limit sample
(4) Test chart DS-001 (LD1625001)
2.1
Acquire White Level Data and Set CIS Scan Area
(1) Wipe the first and second side CIS glass four times in the same direction with a cleaning
cloth.
(2) Press and hold the
(3) Press the 4th blank box. The following screen appears on the LCD.
key. The following screen appears on the LCD.
4-11