Indications
Power/amplitude and mP
have been optimized with no net improvement in PaCO
.
aw
2
Frequency has been lowered to four Hz or less with no improvement in PaCO
.
2
Inspiratory Time % has been increased with no improvement in PaCO
.
2
Procedure
1. With a syringe attached to the endotracheal tube pilot balloon, withdraw air.
2. Reduce the inflation of the cuff until you see a drop in the mP
by
aw
approximately 5 cmH
O.
2
3. Increase the bias flow to correct to the desired mP
level.
aw
Monitoring for cuff leak patency
If mP
increases, suction hypopharynx and reassess. If mP
is lower than desired,
aw
aw
reassess and re-adjust the cuff air pressure to the target mP
level desired. Mean
aw
airway pressure may change after patient position changes; reassessment should be
done after the change.
37