Care Fusion 3100B Pocket Manual page 20

High frequency oscillatory ventilator
Hide thumbs Also See for 3100B:
Table of Contents

Advertisement

If h pH
1.
i Power, maintaining adequate chest wiggle factor.
2.
h Hz.
3.
i % I-time to 33% if at 50%.
If i pH
1.
h Power, obtaining or maintaining adequate chest wiggle factor.
2.
i Hz (minimum of 3.0).
3.
Generate ETT leak.
4.
h % I-time to 50% if at 33%.
hi PaCO
2
1.
Accept hypercapnia if pH allows (> 7.2).
2.
You may consider buffering pH to allow permissive hypercapnia.