High Pressure alarms (alarm setting or > 60 cmH
•
Spontaneously breathing: Consider the clinical status of the patient, assess the
sedation level or insufficient bias flow rate; re-adjust mP
•
Obstruction in the expiratory limb or in the pressure sense line: Replace the
patient circuit
•
Improper setting of the alarm: Change the alarm setting
•
Patient circuit temperature rise: Check and correct the circuit temperature
•
Interference from a radio transmitter: Remove the source of interference
Low Pressure alarms (alarm setting or < 5 cmH
•
Spontaneously breathing: Consider the clinical status of the patient, assess
sedation level or check for insufficient bias flow rate and re-adjust mP
higher flow
•
Improper setting of the alarm: Change the setting
•
Improper setting of the mPAW or flow meter: Change the setting
•
Patient circuit temperature drop: Check and correct the circuit temperature