Managing Oxygenation And Ventilation - Care Fusion 3100B Pocket Manual

High frequency oscillatory ventilator
Hide thumbs Also See for 3100B:
Table of Contents

Advertisement

Managing oxygenation and ventilation

If h PaO
2
1. Wean FiO
slowly (5%) to < 0.60. Re-check the x-ray for lung volume assessment.
2
If lung volume is adequate, continue to wean FiO
approaching a hyperinflation state, consider weaning mP
continue to wean FiO
2. Once FiO
is < 0.40, attempt to wean mP
2
ensuring maintenance of adequate lung inflation and oxygenation.
If i PaO
2
1.
h FiO
as needed to 1.0.
2
2.
h mP
by 3 to 5 every 20 to 30 minutes to obtain adequate lung inflation
aw
and oxygenation.
3.
Check the x-ray to ensure appropriate lung volume.
4.
Check for hemodynamic status for adequate perfusion.
to 0.40.
2
to 0.40. If lung volume is
2
by 1 to 2 every four to six hours,
aw
16
by 1 to 2 cmH
aw
O and
2