Care Fusion 3100B Pocket Manual page 31

High frequency oscillatory ventilator
Hide thumbs Also See for 3100B:
Table of Contents

Advertisement

Troubleshooting equipment issues
Circuit does not pass patient circuit calibration
Visually check for leaks, cracks and open ports on the circuit
Check cap/diaphragms
Check the water trap stopcock (may be open or missing)
Ensure the circuit set up is correctly
Confirm the bias flow is set exactly at 20 LPM (the middle of the ball is at the
20 LPM line—you may need to bend down to see this accurately)
Check the airway pressure luer fittings for cracks
Check the calibration screw (clicking indicates a defective valve)
Confirm the pressure transducer zero: With the circuit stopper in place, but the
system not pressurized, the Paw should read 0 cmH
O (± 0.5 cmH
2
28
O)
2