Fluctuating mP
aw
The Auto-Limit feature is activated
Check the high-pressure setting
Check for spontaneous breathing
Low Source Gas is illuminated
This condition indicates an input pressure of less than 30 psi from either the
blender or the cooling air; check the input gas lines
Ensure that all hoses are plugged into a gas source
Check the blender setup configuration
Some wye or T fittings used to split the high-pressure line have an internal
restriction; remove the wye or T fitting to check it
The input water trap filter needs to be replaced: Replace the filter
Internal leak: Call CareFusion Technical and Clinical Support
30