Weaning Hfov And Transitioning To Conventional Ventilation - Care Fusion 3100B Pocket Manual

High frequency oscillatory ventilator
Hide thumbs Also See for 3100B:
Table of Contents

Advertisement

Weaning HFOV and transitioning to conventional ventilation

When the goals below are met (but no sooner than 24 hours) switch to PCV.
FiO
is weaned to 0.40
2
mP
is 22 to 24 cmH
aw
SpO
> 88% (or as ordered)
2
The patient should be stable on the above settings and able to tolerate
suctioning and brief disconnects. The chest x-ray should show resolution of the
underlying process.
O
2
22