Ventilator does not pass the performance check
Low amplitude:
– Bypass the humidifier
– Check the power knob (0.0 to 10.0)
Low mP
(with or without low amplitude):
aw
– Crimp the airway pressure line (mP
– Check the flow meter
Driver does not start oscillating:
– Check the power knob
– Check mP
– Check if the humidifier chamber is empty (may drop the amplitude by as
much as 10 cmH
– Bypass the humidifier
aw
O)
2
should read 130 to 140 cmH
aw
29
O)
2