Installation and Setup
Power Requirements
Table 1-1
Description
Voltage
Voltage
Power Consumption, On
Power Consumption, Standby
Table 1-2
Description
Voltage
Power Consumption
Power Consumption, Standby

Checking the Fuse

Where IEC regulations apply, use a 5 by 20 mm, rated F5A, 250 V IEC approved fuse. This fuse may
be used with input line voltages of 115 V or 230 V. Its part number is 2110-0709.
Where UL/CSA regulations apply, use a 5 by 20 mm rated fast blow, 5 A, 125 V UL/CSA approved
fuse (part number 2110-0756). This fuse may only be used with an input line voltage of 115 V.
The line fuse is housed in a fuse holder in the upper left hand corner of the rear panel.
To remove the fuse, first disconnect the power cord from the analyzer. Then insert the tip of a
screwdriver into the slot at the middle of the fuse holder, and turn counterclockwise to extend the fuse
holder.
WARNING
For continued protection against fire hazard, replace fuses, and or circuit
breakers only with same type and ratings. The use of other fuses, circuit
breakers or materials is prohibited.
10
AC Power Requirements
90 to 132 Vrms (47 to 440 Hz)
195 to 250 Vrms (47 to 66 Hz)
< 300 W
< 5 W
DC Power Requirements
12 to 20 Vdc
< 200 W
< 5 W
Specifications
Specifications
Chapter 1