Considering Appropriate Installation Conditions - gako unguator EMP Quick Start Manual

Table of Contents

Advertisement

Considering Appropriate Installation Conditions

Assessment
OK
not OK
Stable, flat surface, observe product weight
Suitable space for the device and unobstructed access to the power cord and
power switch
No direct air flow (windows, heating, fans, air conditions)
No direct sunlight
No extreme humidity or temperature fluctuations
Clean, dry, dust-free
Correctly installed, earthed socket with 220 V or country-specific nominal
voltage and required grid frequency
Before initial operation acclimate the device for 30 minutes at room tempera-
ture
Excludes device handling by unauthorized persons
Observe the operation manual and manufacturer's recommendations
chapter 7 - page 2

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for gako unguator EMP

Table of Contents