Common Errors; Cleaning - gako unguator EMP Quick Start Manual

Table of Contents

Advertisement

Common Errors

1
The compound rises up the stirrer shaft.
A suspension ointment is not homogeneous
2
mixed.
The finished compound is at the organoleptic
3
final quality control inhomogeneous.
Cleaning the gako unguator EMP
Always unplug the gako unguator EMP from the wall outlet before performing a cleaning.
Do not use aggressive cleaning agents or abrasive cleaners.
For cleaning, we recommend daily wiping the surface with a damp cloth with mild detergent
and immediately drying it with a dry cloth. For disinfecting, the display can be lightly sprayed
and wiped with isopropanol 70 %.
Make sure that no liquids enter the device. If liquids enter the interior of the gako unguator
EMP, keep the device turned off and inform customer service. An unauthorized opening of the
gako unguator EMP device is not permitted.
Error
tab.: Operation Errors / Common Errors
Troubleshooting
The sealing lip from the lid is damaged.
Replace the lid with an intact one.
The air was not diminished before the mixing
process. Repeat the compounding and
diminish the air correctly before the mixing
process.
The foundation is known to rise (eg. Linola
hydrophilic cream). Refrigerate the foundation
prior the compounding and diminish the air
correctly.
For the mixing process a Disp. Blade was used not
the SMB. Always use the SMB for suspension oint-
ments. The wide blades of the stirrer together with
the inner wall of the jar breaks down agglomerate
and the active substance dispenses homogenic.
Instead of micronized solid substances crystalline
substances are used. To ensure homogeneity use
the ointment mill or micronized solid substances.
,
®
chapter 5 - page 5