Instrument Specifications

The tables found on the following pages contain the technical specifications for the DSC. The specifications
apply to all models, except where noted.

DSC Instrument Characteristics

Dimensions
(Q2000, Q200, Q1000 & Q100)
Dimensions (Auto Q20/Q20/Q10)
Weight (without Autosampler)
Weight (with Autosampler)
Weight of cell alone
Weight of Autosampler alone
Weight of transformer
Power
Accessory Outlets
Operating Environment
Conditions
Temperature Range:
with FACS
with Quench Cooler
with RCS 90
with RCS 40
with LNCS
* You must use an inert atmosphere above 600
** Helium purge gas must be used for temperatures below –100 °C
28
Depth 56 cm (22 in.)
Width 46 cm (18 in.)
Height 48 cm (19 in.)
Depth 56 cm (22 in.)
Width 46 cm (18 in.)
Height 41 cm (16 in.)
25 kg (54 lbs)
29 kg (65 lbs)
1.6 kg (3.5 lbs)
4.5 kg (10 lbs)
8.6 kg (19 lbs)
120 Vac, 47–63 Hz, 500 W (4.5 amps) standard
230 Vac, 47–63 Hz, 500 W (2.25 amps) if configured with a voltage
configuration unit
Power: 120 V, 47–63 Hz, 400 W each
Compatible accessories: LNCS, DPC
Temperature: 15–30 °C
Relative Humidity: 5–80 % (non-condensing)
Installation Category II
Pollution Degree 2
Maximum Altitude: 2000 m
Room temperature to 725
o
-180
C to 550 °C**
-90 to 550 °C
-40 to 400 °C
-180 to 550 °C
o
C.
o
C*
(table continued)
DSC Q Series Getting Started Guide