Q1000/Q100 Autosampler Calibration
To perform Autosampler calibration from the touch screen follow the instructions below.
1.
Touch the AUTOSAMPLER button on the CON-
TROL MENU touch screen. The screen shown in the
figure to the right will be displayed.
2.
Select the desired type of calibration from the buttons
available.
•
To calibrate both the sample tray and cell areas
select AUTO CAL.
•
To calibrate the sample tray only select TRAY
CAL.
•
To calibrate the cell areas only select CELL CAL.
•
If you need to perform more refined calibrations,
select MANUAL CAL to access more calibration
functions.
See the online help for detailed information.
3.
Make sure you save the calibrations when you have finished the desired calibration function.
After calibration is complete, you can perform a Cycle Test to test the calibrated positions.
DSC Q Series Getting Started Guide