TA Instruments DSC Q Series Getting Started Manual page 20

Differential scanning calorimeter
Table of Contents

Advertisement

Control Command
UNLOAD REFERENCE
SWITCH GAS
AIR COOL
PARK AUTOSAMPLER
GO TO STANDBY
TEMPERATURE
LNCS
RCS
RESET SAVED PARAMETERS
SHUTDOWN
20
Description
Unloads the reference pan from the cell to the Autosampler tray
or waste bin.
Toggles between purge Gas #1 and Gas #2.
Toggles the air cool function on or off. This is the same function
as Air Cool on the instrument control software.
Sends the autosampler tray to the park position, which is set off
to the right and below the home position.
When using an RCS or LNCS, this option will cool to the
standby temperature selected in the instrument control program.
Automatically fills the LNCS with liquid nitrogen from the
holding tank when an experiment is not in progress.
Toggles the RCS (Refrigerated Cooling System) between on and
off, depending on the state of the RCS when you press the key.
Resets the saved instrument parameters and resets the instru-
ment.
Shuts down and resets the instrument.
DSC Q Series Getting Started Guide