Dsc Pressure Cell - TA Instruments DSC Q Series Getting Started Manual

Differential scanning calorimeter
Table of Contents

Advertisement

DSC Pressure Cell

The Pressure DSC (PDSC) cell is a DSC cell enclosed in a steel
cylinder that can be pressurized to 7 MPa (1000 psig). In addition
to performing the same measurements as the DSC cells, it can
operate at elevated pressure or under vacuum. This ability to
vary pressure as well as temperature provides the following:
Resolution of overlapping peaks
Determination of heats of vaporization and vapor pressure
Reaction rates in controlled atmospheres
Studies of pressure-sensitive reactions
The Pressure DSC cell has two gas flow control valves, a three-
way valve, a pressure gauge, a pressure release valve, and gas
pressure fittings on the side. The cell has a pressure transducer
which continuously measures actual sample pressure and stores
it in the data file.
This cell is optional for the Q2000 and Q1000. It comes standard
on the Q20P and Q10P.
DSC Q Series Getting Started Guide
27

Advertisement

Table of Contents
loading

This manual is also suitable for:

Q10Q1000Q20Q2000Q100Q200

Table of Contents