Section 7: Troubleshooting; General; Symptom, Cause, And Remedy - Dräger Medical Isolette C2000 Operating Instructions Manual

Infant incubator
Table of Contents

Advertisement

General

WARNING:
Only facility-authorized personnel should perform preventive maintenance or troubleshooting
on the Isolette® Infant Incubator, Models C2000 and C2000e. Preventive maintenance
performed by unauthorized personnel could result in personal injury or equipment damage.
If the fault cannot be located, refer the unit to qualified service personnel.

Symptom, Cause, and Remedy

For troubleshooting of the incubator, refer to the table below:
Symptom
There is no system power, and the
Power Fail alarm does not acti-
vate.
The Power Fail alarm activates.
The Low Air /Skin Temperature
alarm activates.
Low oxygen concentrations exist
Troubleshooting
Possible Cause
The incubator Main Power switch
is Off.
For non-UPS systems, only:
The power cord is unplugged.
For non-UPS systems, only:
The power cord to the incubator is
unplugged.
For UPS systems, only:
The batteries are not charged.
An access door or iris entry port is
open.
The skin probe is not properly
secured to skin (in Skin mode
only).
An access door or iris entry port is
open.
Section 7
Troubleshooting
At the stand, turn on the Main
Power switch.
Make sure that the power cord is
plugged into the AC power source.
If the power cord is detachable,
also make sure it is firmly attached
to the stand receptacle.
Re-attach the power cord to the
incubator receptacle.
After a power failure or prior to
initial usage, keep power applied
continuously to the UPS system
for a maximum of 8 hours.
NOTE:
Ensure the Resettable circuit
breaker is in the On (1)
position.
Close all access doors and iris
entry ports.
Check the skin probe connection.
Close all access doors and iris
entry ports.
Remedy
7 - 1