CEM EDGE system Manual page 52

Energized dispersive guided extraction
Hide thumbs Also See for EDGE system:
Table of Contents

Advertisement

4. Select Temperature "Calibrate."
5. Select "OK" to begin temperature calibration. The actuator will close, the dispense needle will travel to waste
and the system will begin filling the chamber with water. Follow the remaining software prompts to complete
the calibration.
5.1. After 4:00 min, you will be prompted to take the temperature of the water; be sure to mix the water.
5.2. After 0:20 seconds of mixing, the software will prompt you to type in the final temperature you are reading.
Type in the value, and select "OK."
5.3. Verify temperature reading, and select "OK." The system will begin flushing the water into waste and will
then return to the home position.
6. Once calibration is complete select "OK."
48
Calibrations