4. Select Pressure "Calibrate." This will take less than a minute.
5. Select "OK."

Verify Autosampler Calibration

The autosampler requires calibration verification during the initial install, if the EDGE system is moved, and semi-
annually.
1. Select the System Menu icon
2. Select Tools.
3. Scroll down to the bottom of the page to the Autosampler section.
CAUTION
When verifying the autosampler calibration, always select "Load," "Collect", and "Unload" in that order
without skipping a function.
4. Place a Q-Cup and collection vial in position 4 of the rack.
NOTE
Position 4 is used to verify the calibration. Moving the Q-Cup to and from this location verifies the
interpolation of the calibration positions and is therefore the best test for the calibration.
5. Select "1" from Sample Position to display the sample position menu, and select position 4.
in the bottom right corner of the screen.
Calibrations
43