NOTE
Waste bottle must be attached before priming solvent lines.
10. Select the lines you would like to prime, and press "OK." The selected bottle position lines will be primed.
11. Once complete and a green check mark is shown, select "Close."
12. If a green check mark is not shown, there is a possible leak in the solvent line. Please contact CEM Corpora-
tion.
Remove a Solvent from Bottle Setup Position
1. Select System Menu icon
2. Select Settings.
3. Select Bottle Setup.
4. Select the desired bottle position.
5. Select the solvent to be removed.
6. Select "OK."
7. Select the Save icon