Opening the LabChip GX Touch Software

To open the LabChip GX Touch software:
1
2
3

Purging the Pressure Lines

The LabChip GX Touch software provides a Purge Pressure Lines
function to improve instrument performance. A positive pressure is
applied to the lines and removes any potential liquid or debris out of
the lines through the chip interface.
Purge the pressure lines at the start of each day, before inserting a
chip.
To purge the pressure lines:
1
2
P/N CLS151164 Rev. B
Touch the LabChip GX Touch icon on the Windows desktop.
If the LabChip GxP option is installed, the
Window
opens. Type a valid LabChip GX Touch user name and
password into the text boxes and touch the Logon button.
LabChip GX Touch user names are created in the LabChip GX
Reviewer software. (For instructions on creating LabChip GX
Touch user names, see the LabChip GX Reviewer User
Manual.)
The
LabChip GX Touch Main Window
Bar
displays on the left side of the LabChip GX Touch Main
Window, and the
Home Window
If a chip is already inserted into the instrument, remove the chip
and close the chip door (see
Touch the Purge Pressure Lines button on the
The
Status Window
opens. The purge is complete when the Run
Status displays "Purge successfully completed" and the
instrument status returns to Idle.
LabChip GX Touch/GXII Touch User Manual
Preparing the Instrument to Run Samples
opens. The
displays in the center.
page
41).
24
Unlock Application
Navigation
Home
Window.
PerkinElmer, Inc.