4.10 Fine Span Adjustment

F span
Fine Span Adjustment (from SETUP Prompt)
Fspan?
Fine Span Adjustment (after pressing TARE  at last Print prompt)
8400-M022-O1 Rev J
IMPORTANT! The FSPAn (FSPAn?) mode requires a
load of 10% of Capacity on the scale before adjustments
can be made.
If Fine Span Adjustment is desired, with the SETUP prompt
1.
displayed, press the UNITS/ key until the display shows the
FSPAn prompt.
Place a calibrated test weight on the scale and press the TARE  key.
2.
The display will change to show the amount of the test weight, an F
3.
will be displayed to the right of the annunciators and the
annunciators will alternately flash off and on.
Press the Fn/ key to increase the span by 0.5 division OR press
4.
the UNITS/ key to decrease the span by 0.5 division.
Press the TARE  key to exit the Fine Span Adjustment and
5.
advance to the Hires? prompt.
With fspan? displayed, press the TARE  key.
1.
The display will change to no.
2.
Place a calibrated test weight on the scale.
3.
Press the Fn/ key to toggle to yes and then press the TARE  key.
4.
The display will change to show the amount of the test weight, an F will
5.
be displayed to the right of the annunciators and the annunciators will
alternately flash off and on.
Press the Fn/ key to increase the span by 0.5 division OR press the
6.
UNITS/ key to decrease the span by 0.5 division.
Press the TARE  key to exit the Fine Span Adjustment and advance
7.
to the Hires? prompt.
Model 190 Installation, Technical and Operation
Page 61