4.4 Settings

SETUP
USA= (Domestic or International)
LFt= (Legal For Trade)
8400-M022-O1 Rev J
IMPORTANT! Calibration and configuration parameters are not
stored in the non-volatile memory until setup is left. If power is
lost while in setup mode, any changes made will be lost and the
indicator will revert to the previous configuration.
With SEtUP displayed, press the TARE  key. The display will change
to USA=. Press the TARE  key to show the current setting. If the
setting displayed is acceptable, press the TARE  key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and then press
the TARE  key to save it.
Yes (Domestic)
trL= (4% Zero Range) = no
CAP (Capacity) = + 4% to OC
Date Format = MM/DD/YY
Press the TARE  key to show the current setting. If the setting
displayed is acceptable, press the TARE  key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and then press
the TARE  key to save it.
Yes = Interval Settings (Int=)
allowed are: 1, 2, 5, 10, 20, 50
When both LFt=yes and USA=yes, the followings results occur:
Scale must have between 100 and 10,000 divisions
Inhibit serial data during input
trA= (Zero Tracking Range) = 0.5 or 0 to 3
trL= (4% Zero Range) = no
CAP (Capacity) = + 4% to OC
When LFt=yes and USA=no, the followings results occur:
UnS= (Motion Range) = 1
trL= (4% Zero Range) = yes
CAP (Capacity) = + 9 grads to OC
Model 190 Installation, Technical and Operation
no (International)
trL=(4% Zero Range) = yes
CAP (Capacity) = + 9 grads to OC
Date Format = DD/MM/YY
no = Interval Setting (Int=) is
selectable from 1 to 99.
Page 23