Guardrails® Editor Software Profile Configuration (Continued)
The following configurations are only used when the Alaris® CC Guardrails® Syringe Pump is being used in ml/h or Dosing Only modes. (If a
drug is selected then the drug's own configuration settings are used.)
Bolus
Default Bolus Rate
Max Bolus Volume
Occlusion Alarm Pressure
Units Enabled for Dosing Only
The approved Data Set contains configurable option values per profile.
w
The originator and approvers of the Data Set should be aware that, unless a rationale for safety is provided, it is not
recommended to set the callback time to a value greater than the default setting of 2 minutes since doing so would not
be in compliance with EN60601-2-24:1998 standard.
Guardrails® Editor Software Profile Drug library
The following drug parameters are only configurable via the Guardrails® Editor Software (PC Based), see Guardrails® Editor Directions For
Use (1000PB01398) for details on how to configure Profile Drug Library, and are used when the Alaris® CC Guardrails® Syringe Pump is
being used with a drug name selected.
Concentration Limits (Min & Max)
Continuous Dose Rate -
Units
Hard Limit Max
Soft Alert Max
Soft Alert Min
Default
Bolus
Bolus Dose -
Units
Hard Limit Max
Soft Alert Max (HANDS FREE only)
Soft Alert Min (HANDS FREE only)
Default (HANDS FREE only)
Bolus Rate -
Default
Occlusion Alarm Pressure
Configured Options (Continued)
Bolus feature can be set to OFF, HANDS ON or HANDS ON & HANDS FREE.
The default value for bolus rates.
The maximum permissible bolus volume.
The default occlusion pressure alarm level.
Dosing only mode on an Alaris® Syringe Pump allows the pump to be set-up with a dosing protocol
with no drug selected. Only the dosing units selected in Units Enabled for Dosing Only will be
available for selection when the Alaris® Syringe Pump is in Dosing Only mode.
These define the range over which the drug concentration can be modified during
programming of the Alaris® CC Guardrails® Syringe Pump.
The continuous dose rate units. Can be based on patient weight.
The maximum allowed continuous dose rate.
The continuous dose rate value above which override confirmation is required.
The continuous dose rate value below which override confirmation is required.
The default continuous dose rate offered when the drug is selected.
Bolus feature can be set to OFF, HANDS ON or HANDS ON & HANDS FREE.
The bolus dose units. Can be based on patient weight.
The maximum allowed bolus dose.
The bolus dose value above which override confirmation is required.
The bolus dose value below which override confirmation is required.
The default bolus dose offered.
The default value for bolus rate in ml/h.
The default occlusion alarm pressure in mmHg.
1000DF00333 Issue 6