CareFusion Alaris CC Guardrails 80033UND1-G Directions For Use Manual page 19

Syringe pump
Table of Contents

Advertisement

This option allows you to set a specific volume to be infused. Rate at the end of this VTBI can also be set, selecting from stop, KVO, or
continuous infusion at the set rate.
1. Press the VTBI softkey to select the volume to be infused option.
2. Enter the volume to be infused using the
3. Select the rate at the end of the VTBI using the
4. Press the OK softkey to enter the rate and exit the VTBI menu.
This option enables the volume infused to be cleared.
1. Press the VOLUME softkey to display the CLEAR VOLUME option.
2. Press the YES softkey to clear the volume. Press the NO softkey to retain the volume.
Selecting YES resets the volume infused in the 24H LOG option.
If Rate Lock is enabled, when the infusion rate has been set and the infusion started (or following a bolus infusion) the rate lock prompt
will appear on the main display.
To select the rate lock function press the YES softkey. Press the NO softkey if the rate lock is not required.
When rate lock is enabled, the following are unavailable:
l Changing the infusion rate / titration
l Bolus / purge
l Switching the pump off
l VTBI over time infusions.
To disable the rate lock if selected:
1. Press the
d
button to access the options menu.
2. Select the UNLOCK RATE option using the
To enable the rate lock if not selected:
1. Press the
d
button to access the options menu.
2. Select RATE LOCK and press the OK softkey.
If Rate Titration is enabled the rate can be adjusted while infusing:
1. Select the new rate using the
The message < START TO CONFIRM > will flash on screen and pump continues to infuse at the original rate.
2. Press the
b
button to confirm the new infusion rate and start infusing at the new rate. If the new infusion rate setting exceeds or is
under a Guardrails® Soft Alert confirmation is required before infusion can start infusing at the new rate.
If Rate Titration is disabled the rate can only be adjusted whilst on hold:
1. Press the
h
button to put the pump on hold.
2. Select the new rate using the
3. Press the
b
button to start infusing at the new rate.
Basic Features (Continued)
Volume to be Infused (VTBI)
f
keys and press the OK softkey.
f
f
keys and press the OK softkey.
f
keys.
f
keys.
1000DF00333 Issue 6
keys to scroll through the on-screen choices. The default is stop.
Clear Volume
Rate Lock
Rate Titration
18/38