8. CONFIRM SYRINGE - Check that the syringe type and size being used matches the display. If required, the make of syringe can be
changed by pressing the TYPE button. Press CONFIRM when the correct type and size are shown.
Note: If the PURGE SYRINGE option has been enabled then the prompt to purge screen is displayed and the extension set can be
purged as required.
9. PURGE (if required) - Press the
set is complete. Release the softkey. The volume used during purging will be displayed.
Purge extension set, massaging pressure disc to prevent ballooning and ensuring all air removal.
w
10. INFUSION RATE - Check the rate shown if set and change the rate if necessary using the
11. CONNECT TO PATIENT - Connect the extension set to the patient access device.
12. START - Press
b
to commence operation. INFUSING will be displayed. If the infusion settings are within the Guardrails® Soft Alerts then
the AMBER STOP light will be replaced by the flashing GREEN START light to indicate that the pump is in operation.
If infusion rate exceeds the Guardrails® Hard Limit then the pump will not start and the display will show DOSE NOT PERMITTED.
If infusion rate exceeds or is under the Guardrails® Soft Alerts then check infusion setting, to continue with infusion at set rate press
and then confirm OVERRIDE LIMIT by pressing YES. If OVERRIDE LIMIT is not required press NO and adjust rate to be within the
b
Guardrails® Soft Alerts.
If infusion rate running exceeds or is under the Guardrails® Soft Alerts then the display will cycle between Drug
w
Name, Profile name and Up or Down arrows.
13. STOP - Press
h
to halt the operation. ON HOLD will be displayed. The AMBER STOP light will replace the GREEN START light.
FULLY DEDICATED Alaris® CC Guardrails® Syringe Pump - to start an infusion a pressure disc must be fitted.
SEMI DEDICATED Alaris® CC Guardrails® Syringe Pump - to start an infusion with Drug Name or Dosing Only selected a pressure disc
must be fitted.
Starting the Pump (Continued)
i
button and then press and hold the PURGE softkey until fluid flows and the purging of the extension
1000DF00333 Issue 6
f
keys.
16/38