The following options are only configurable via the Guardrails® Editor Software (PC based), see Guardrails® Editor Directions For Use
(1000PB01398) for details on how to configure Profile Configurations.
Quiet Mode
Audio Volume
Auto Night Mode
AC Fail
Auto Save
Event Log
Battery Icon
Callback Time
Drug Override Mode
Maximum Pressure
Auto Pressure
Auto Set Pressure
Auto Offset
Back Off
Pressure Display
Default Weight
Maximum Weight
Minimum Weight
Rate Lock
Rate Titration
Cap Rate
Cap Bolus Rate
Purge Rate
Purge Volume Limit
Purge Syringe
Manual Bolus
KVO
Near End of Infusion Point
End of Infusion %
VTBI Clear Rate
Configured Options (Continued)
Guardrails® Editor Software Profile Configuration
Mode to silence key press tones and power down sequence.
The audio alarm volume of the pump (high, medium or low).
Main Display (Backlight) dims between hours 21:00 and 06:00.
The AC Power Failure Alarm can be set to sound or be silent if the AC power is
disconnected.
Feature to retain previous settings when pump is switched on.
The event log can be set to be displayed or not on main display. Events still recorded in
event log if disabled.
Indicator displaying the remaining estimated battery capacity.
Adjusts the length of time before the pump sounds the call back alarm.
Always - Any changes made to the dose rate that are outside of the Guardrails® Soft Alerts
will require confirmation before starting infusion.
Smart - Confirmation of setting will be required on first dose rate set outside of the Guardrails®
Soft Alert. Any subsequent changes will not require confirmation until after the dose rate has been
confirmed inside the Guardrails® Soft Alert limits. Additionally any changes in dose rate from above
a Soft Alert Max to below a Soft Alert Min or from below a Soft Alert Min to above a Soft Alert Max
will also need to be confirmed.
The maximum occlusion pressure alarm level that can be selected during an infusion.
Feature to set the occlusion pressure alarm level to an amount (mmHg) above the current in-line
pressure, using a single key press.
Automatic feature to set the occlusion pressure alarm level to an amount (mmHg) above the current
in-line pressure, 15 minutes after starting the infusion.
The automatic offset value in mmHg used by auto pressure and auto set pressure.
An automatic feature which is activated following an occlusion. The pump action reverses and
pumps backwards to release the pressure which has built up in the infusion system, this minimises
the post occlusion bolus.
Sets whether the Pressure Information is available on the main display.
The default patient weight in kg.
The maximum patient weight in kg. This is a Guardrails® Soft Alert and can be overridden.
The minimum patient weight in kg. This is a Guardrails® Soft Alert and can be overridden.
Anti-tamper feature which prevents rate changes, bolus operations and powering pump down.
Feature to adjust the infusion rate while the pump is infusing, without putting the pump on hold
The maximum value for infusion rate.
The maximum value for bolus rate.
The rate used during purge operation.
The maximum permissible purge volume.
Feature which prompts the user to purge the extension set prior to the start of the infusion.
Bolus delivered by manually moving the plunger mechanism during an infusion or while on hold.
Volume infused displayed will be increased accordingly.
Sets the Keep Vein Open (KVO) rate at which the pump will operate when End of Infusion (EOI) is
reached. Also allows KVO at EOI to be disabled.
Sets the Near End Of Infusion warning time, as time left to End Of Infusion.
Sets the End Of Infusion point, as a percentage of syringe volume.
Infusion rate will be set to zero when VTBI has been completed.