CareFusion Alaris CC Guardrails 80033UND1-G Directions For Use Manual page 20

Syringe pump
Table of Contents

Advertisement

If the Auto Set Pressure Option is enabled then the pump AUTOMATICALLY adjusts the pressure occlusion limit.
After 15 minutes of infusion the pump AUTOMATICALLY adjusts the occlusion pressure limit to X mmHg (the AUTO OFFSET value, 15-
l
100mmHg), above the average infusion pressure since the start of the infusion.
1. To check and adjust the pressure level press the
the pressure alarm level and the current pressure level.
2. Press the
f
keys to increase or decrease the pressure alarm level. The new level will be indicated on the display.
3. The AUTO Pressure feature may be used when a stable pressure has been achieved over a short period of infusion. If AUTO Pressure
has been enabled the automatic pressure alarm level is calculated and set by pressing the AUTO softkey.
4. Press the TREND softkey to view the pressure trend of the previous 12 hours. The pressure trend can be viewed at 15 minute intervals
by using the +/- softkeys. The pressure trend graph displays the pressure at a given time.
5. Press the OK softkey to exit the pressure screen.
1. To check and adjust the pressure level press the
current pressure level.
2. Press the
f
keys to increase or decrease the alarm level. The new level will be indicated on the display.
3. Press OK to exit the screen.
The interpretation of pressure readings and occlusion alarms are the responsibility of the clinician depending on the
w
specific application.
Basic Features (Continued)
Auto Set Pressure (If enabled)
Pressure Level with pressure set fitted
e
button. The display will change to show a 20 minute pressure trend graph displaying
20mmHg
Pressure Level without pressure set fitted
(not applicable when FULLY DEDICATED)
e
button. A bar graph will be displayed showing the pressure alarm level and the
1000DF00333 Issue 6
07:45
_
+
OK
19/38