Controls & Indicators - CareFusion Alaris CC Guardrails 80033UND1-G Directions For Use Manual

Syringe pump
Table of Contents

Advertisement

Controls:
Symbol
a
b
h
c
i
d
e
f
g
Indicators:
Symbol
j
S
Controls & Indicators
Description
ON/OFF button - Press once to switch the pump ON. Press and hold down for 3
seconds to switch the pump OFF.
RUN button - Press to start the infusion. The green LED will flash during infusion.
HOLD button - Press to put the infusion on hold. The amber LED will be lit while on
hold.
MUTE button - Press to silence alarm for 2 minutes (configurable). Press and hold
until 3 beeps are heard for 15 minutes silence.
PURGE/BOLUS button - Press to access PURGE or BOLUS soft keys. Press and hold
down soft key to operate.
PURGE the extension set during set up.
Pump is on hold
l
Extension set is not connected to the patient
l
Volume Infused (VI) is not added
l
BOLUS - fluid or drug delivered at an accelerated rate.
Pump is infusing
l
Extension set is connected to the patient
l
VI is added
l
OPTION button - Press to access optional features (see Basic Features).
PRESSURE button -
alarm level.
CHEVRON keys - Double or single for faster/slower increase or decrease of values
shown on display.
BLANK SOFTKEYS - Use in conjunction with the prompts shown on the display.
Description
BATTERY indicator - When illuminated the pump is running on the internal battery.
When flashing the battery power is low with less than 30 minutes of use remaining.
AC POWER indicator - When illuminated the pump is connected to an AC power
supply and the battery is being charged.
1000DF00333 Issue 6
Use this button to display the pumping pressure trend display and
5/38