The Parameters

FSC:
forward scatter
SSC:
side scatter
FL1:
green fluorescence
fluorescence origin: 488 nm laser (GFP)
FL2:
orange fluorescence
fluorescence origin: 488 nm laser (PE)
FL3:
blue fluorescence
fluorescence origin: UV-LED (DAPI)
FL4:
red fluorescence
fluorescence origin: 638 nm laser (APC)
Example: This instrument is equipped with a blue
diode pumped solid-state Laser (20 mW) at 488
nm, a red laser diode (25mW) at 638 nm and a
365 nm UV-LED.
Rev018_2012-01-27
Optical bench of a CyFlow
32/36
If required by a specific application, the optical
standard setup can be optimized by exchanging
preassembled removable mirror/filter blocks. This
is a matter of seconds and does not require any
re-adjustment.
®
The CyFlow
Cube flow cytometer can be
equipped with various light sources and up to
eight optical parameters. Depending on the
number of light sources and optical parameters
different optical benches are available.
Due to its modular concept the optical
configuration can be adopted to many different
clinical and scientific purposes. Standard
configurations are presented below:
®
Cube with 2 lasers, UV-LED and 6 parameters.