Instrument Starting Procedure; Sheath Fluid Level Control And/Or Refill; Switching On The Cyflow ® Cube - Partec CyFlow Cube Series Instrument Operating Manual

Table of Contents

Advertisement

Instrument starting procedure

Sheath fluid level control and/or refill

Before
switching
recommended to check the levels of the sheath
fluid and waste bottles. They can be found at the
back-left
of
the
compartment.
Make sure SHEATH bottle is filled with 800 ml of
clean, filtered, and degassed sheath fluid and is
closed with the screw top. In order to guarantee
highest quality of the measurements we highly
recommend use Partec Sheath Fluid (order no.
04-4007).
It is recommended to replace the sheath fluid at
least once a week or before any daily use.
When filling up the sheath fluid bottle make sure
no air bubbles are trapped in the yellow filter unit
inside the bottle!
Make sure WASTE bottle is empty and the screw
top is tightly closed.
The waste bottle must be emptied after and
before each user session. When using bio-
hazardeous samples, a volume of 50 ml of
hypochlorite 0.5% (Order No 04-4012) should be
introduce into the empty waste bottle for initial
disinfection.
Switching on the CyFlow
The power supply switch is found at the back of
the Cube next to the main supply cable. The Cube
is, in default, set in a stand-by mode. The full
activation of the Cube requires pressing lightly the
on/off tactile button on the top of the machine.
The display screen must be first lift up to access
it.
This
will
start
the
automatically start the CyView
the last employed configuration.
on
the
machine,
apparatus
in
a
sliding
®
Cube
embedded
computer,
software and load
CyFlow
it
is
Main
power
switch
Rear side of CyFlow
power switch and the sheath fluid and waste
bottle compartment.
Casual/medium expertise user:
Once the Cube started no further steps are
necessary. You can directly start your
11/36
®
Cube Instrument Operating Manual
Waste and
sheath
fluid
®
Cube showing the main
measurements!
On/Standby
Rev018_2012-01-27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cyflow cube 6Cyflow cube 8Cyflow cube 8 sorter

Table of Contents