Fast Condensation Particle Counter Model 3650: Installation
5. Connect the bypass inlet to an ion mobility classifier with conditioner
or stainless steel tubing.

Connecting the n-Butanol Supply

The fill bottle and bottle bracket are supplied with the Fast CPC. To mount
the n-Butanol bracket and fill bottle, follow these instructions:
1. Using a Phillips screwdriver, remove the two screws on the back
panel of the Fast CPC (beneath the fan).
2. Using the screws, attach the bottle bracket to the back of the Fast CPC
in the position shown below.
3. Snap the provided fill bottle into place in the bracket as shown in the
photo below.
4. Insert the fitting on the end of the bottle's lower tube into the n-
Butanol Supply inlet.
5. Insert the fitting on the end of the bottle's upper tube into the n-
Butanol Exhaust outlet.