Priming the Flow

The Prime button allows you to increase the butanol injection rate by 10x to
reduce the time needed to fill the butanol pumps and tubing. When Prime is
active the injection rate increases by 10x.
To prime the flow, follow these instructions:
1. Press F2.
2. On the Device Settings screen, touch Flow Settings.
3. On the Flow Settings screen, touch the Prime toggle buttons to turn the
priming function On.
4. When Butanol is consistently exhausted from the instrument, press Prime
On to turn Off the Prime function. Note: It is normal for bubbles to be
present in the Butanol exhaust tubing. Excessive priming may cause
flooding in the inlet.