Fast Condensation Particle Counter Model 3650: Installation

How to Install the Fast CPC Model 3650

Following is an overview of the steps required to install a Fast CPC. Please
read the detailed instructions (beginning below) for each step before you set
up the instrument.

Unpacking the Fast CPC.

Connecting the aerosol supply.
Connecting the n-Butanol supply.
Connecting the power.
Installing/removing the inlet bypass fitting.
Connecting the bypass outlet.
Unpacking the Fast CPC
To unpack the Fast CPC, follow these instructions:
1. Carefully remove the Fast CPC from its shipping container. Save the
original packing materials for use when shipping the Fast CPC back to
Kanomax FMT, Inc. for service, or for moving the Fast CPC to a different
location.
Warning. If the Fast CPC is returned to Kanomax FMT, Inc. in anything
other than the original shipping container, you will be charged for any
damage that occurs during shipping. If you do not have the original
shipping container, contact Kanomax FMT Inc. at 651-762-7762.
(Customers in Asia please call +81 6-6877-0183.)
2. Place the Fast CPC on a level surface.
3. Make sure there is an unrestricted air flow around the device. Kanomax
FMT, Inc. recommends at least a 2-inch air gap on both sides and the top
of the instrument.
4. Allow the Fast CPC to reach ambient temperature, if necessary.
5. Make sure all the items listed in Table 1, were included in the Fast CPC
shipment. If any of the items are missing, or damaged, please call
Kanomax FMT, Inc. at 651-762-7762. (Customers in Asia please call +81 6-
6877-0183.)
Fast CPC User Manual
7
Version 1.2