Warranted Characteristics
1 - 24
Table
1-13:
Warranted Characteristics-
Probe Adjust Output
Name
PROBE ADJUST Accuracy
Description
Measurement
Output Voltage
Repetition Rate
Accuracy
0.5 V±5%
1 kHz±20%
Table
1-14:
Warranted Characteristics-
Power Requirements
Name
Line Voltage Ranges
Line Frequency Range
Max. Power Consumption
Line Fuse
Description
95 VAC to 128 VAC or 190-250 VAC
( depending on Line Voltage Setting)
a
48 Hz to 440 Hz
85 Watts (95 VA)
Line Voltage
Range
Fuse
95-128 VAC Range 1 A, slow blow, 230 V
190-250 VAC Range 0.5 A, slow blow, 230 V
To change the Line Voltage Range inside the instrument, a qualified technician
must be consulted to change the fuse and the power setting.