Tektronix 2212 Service Manual page 105

Digital storage & analog oscilloscope
Hide thumbs Also See for 2212:
Table of Contents

Advertisement

Brief Procedures
Table
4-1:
Test Equipment Required
Item and
Minimum
Example of
Description
Specification
Use
Test Equipment
1. Calibration
Standard-amplitude sig-
Signal source
Tektronix
Generator
nal levels: 5 mV to 50 V.
for gain and
PG 506A
Accuracy: ±0.3 %.
transient
Calibration
- High-amplitude signal
response_
Generator
a
levels: 1 V to 60 V.
checks ~d
Repetition rate: 1 kHz.
adjustments
- Fast rise signal level:
1 V. Repetition rate:
1 MHz. Risetime: 1 ns
or less. Flatness: ±0.5%.
2. Leveled
Frequency: 250 kHz to
Vertical,
Tektronix
Sine-Wave
above 60 MHz.
horizontal, and
SG503 Leveled
amplitude
Output:: variable from
triggering
Sine-Wave
Generator
1 O mV to 5V p-p. Out-
checks and
Generator
a
put impedance: 50
n.
adjustments.
Reference frequency:
Display adjust-
50 kHz. Amplitude
ments and
accuracy: constant with-
Z-Axis check
in 3% of reference
frequency as output
frequency changes.
3. Time-Mark
Marker outputs: 1 O ns to
Horizontal
Tektronix
Generator
0.5 s. Marker accuracy:
checks and
TG501 Time-
±0.1 %. Trigger output:
adjustments.
Mark Genera-
1 ms to 0.1
µs,
time-
Display
tor•
coincident with
adjustment
markers.
4.
Low-
Range 1 O Hz to
Low-
Tektronix
Frequency
500 kHz.Output ampli-
Frequency
SG502
Sine-Wave
tude: 300 mV. Output
trigger
Oscillato
a
Generator
impedance: 600
n.
Ref-
checks
erence frequency:
constant within 0.3 dB
of reference frequency
as output frequency
changes.
5. Pulse
Repetition rate:
Signal source
Tektronix
Generator
1 kHz. Output
for Storage
PG501
amplitude: 5 V.
and external
Pulse Genera-
clock checks
tor
a
2212 Service Manual
4-3