Cursors Adjustments; Adjust Cursors - Tektronix 2212 Service Manual

Digital storage & analog oscilloscope
Hide thumbs Also See for 2212:
Table of Contents

Advertisement

Adjustment Procedures
Cursors Adjustments
5- 20
Equipment Required
None
Initial Control Settings
Calibration MODE
Adjust Cursors
Procedure
Calibration via the Calibration Routine in the 2212
a. Place: CAL jumper in CAL position, and wait for the
Calibration MENU on the CRT screen.
b. Select: CURSORS to GO.
c. Adjust: Cursors by following the instructions at the screen.
d. Remove: CAL jumper to enter normal scope operation.
Adjustment