Brief Procedures; Initial Setup Procedure - Tektronix 2212 Service Manual

Digital storage & analog oscilloscope
Hide thumbs Also See for 2212:
Table of Contents

Advertisement

Initial Setup
Procedure
Brief Procedures
This subsection contains a collection of procedures for adjusting the
2212
Analog
&
Digital Storage Oscilloscope to meet the Performance Requirements.
Adjustments described are:
Power supply and CRT Adjustments
Vertical Adjustments
Horizontal Adjustments
Trigger Adjustments
Cursors Adjustments
Adjustments contained in this section should only be performed after checks
from Section
4
(Performance Check Procedure ) have indicated a need for
readjustment, or after instrument repair.
Once an adjustment has been made, use the Performance Check Procedure to
check the performance of the instrument.
This procedure set.,; the front-panel controls for the tests that follow.
Procedure
Stepl.
Plug the female connector of the power cord in the power cord recep-
tacle of the
2212
and the male connector to the AC power source.
Step 2.
Use the test equipment, as indicated in the Equipment Required list, to
the
2212
oscilloscope.
Step 2.
Press the POWER button to on.
Step 3.
Set the front-panel controls as indicated in the Initial Control Settings
list.
2212
Service Manual