PERFORMANCE CHECKS
Description
Check
1.
CIRCUIT CHECK Meter
2.
5 MHz, 1 MHz, 100 kHz Outputs
3.
Output Voltages/Waveforms
4.
Harmonic Distortion
5.
Non-Harmonic Distortion
6.
Frequency Stability
7.
Option 001 Time Standard
RMS Deviation:
1 sec averaging
10 sec averaging
100 sec averaging
See Table 5-3
Correct Frequency
at least 1 V rms.
40 dB
80 dB
1x10-
11
Rate: 1PPS
Width: 20/isec
Level: +10Vp-p+10%
Rise Time: 50nsec
Fall Time: 1/jsec
Jitter: 1 nsec
Delay: 10/usecto 1 sec
I
I
* v
I
I
1fi
I
I
fiy
10"
12
x 1 0
1 2
10-
13
5-17