6 Set up the sensor with the display and adjustment module
Display - Displayed value
1 and 2
Display - Display format
1 and 2
Display - Backlight
Diagnostics - Device
status
Diagnostics - Peak val-
ues, pressure
32
•
Chinese
•
Polish
•
Czech
•
Turkish
In delivery status, the VEGABAR 86 is set to English.
In this menu item, you define which measured value is displayed.
The setting in the delivery status for the display value is "Lin. percent".
In this menu item you define the number of decimal positions with
which the measured value is displayed.
The setting in the delivery status for the display format is "Automatic".
The display and adjustment module has a backlight for the display.
In this menu item you can switch on the lighting. You can find the
required operating voltage in chapter "Technical data".
In delivery status, the lighting is switched on.
In this menu item, the device status is displayed.
In case of error, e.g. the error code F017, e.g. the error description
"Adjustment span too small" and a four digit figure are displayed for
service purposes. You can find the error codes with description, rea-
son as well as rectification in chapter "Asset Management".
The respective min. and max. measured values are saved in the
sensor. The two values are displayed in menu item "Peak values,
pressure".
In another window you can carry out a reset of the peak values
separately.
VEGABAR 86 • Modbus and Levelmaster protocol