System Temperature Error For The Dt9871 - Data Translation DT9871 User Manual

Standard temppoint, voltpoint, and measurpoint usb instruments
Hide thumbs Also See for DT9871:
Table of Contents

Advertisement

Appendix A

System Temperature Error for the DT9871

Table 11
temperature points over the dynamic range of the instrument.
Input
Temperature
–100C
0C
100C
300C
500C
700C
900C
1100C
1400C
a. Conditions for accuracy measurements:
MEASURpoint warm-up time of 1 hour. Inclusive of typical 0.2
Exclusive of thermocouple errors.
The histograms shown in
distribution for each of the available filter types on the DT9871. Note that converting V error
to temperature error depends on thermocouple type. For example, a K thermocouple changes
approximately 39 V per degrees C; therefore, a noise level of 10 V adds 0.3
(10 V / 39 V) for a type K thermocouple.
114
lists the typical accuracy of the DT9871 for each thermocouple type at several
Table 11: Typical Thermocouple Accuracy of the DT9871
J
K
T
0.33C
0.37C
0.38C
0.30C
0.33C
0.33C
0.31C
0.29C
0.31C
0.29C
0.34C
0.29C
0.28C
0.31C
-----------
0.27C
0.30C
-----------
0.27C
0.34C
-----------
0.28C
0.34C
-----------
-----------
-----------
-----------
Figure 38
a
Thermocouple Type
E
S
0.31C
-----------
0.28C
1.12C
0.27C
0.88C
0.26C
0.75C
0.26C
0.71C
0.27C
0.68C
0.28C
0.66C
0.64C
-----------
0.62C
-----------
C CJC error (maximum CJC error is 0.3
and
Figure 39
characterize the Gaussian system noise
R
B
0.44C
-----------
-----------
1.14C
0.39C
-----------
0.86C
0.35C
-----------
0.71C
1.84C
0.33C
0.66C
1.20C
0.33C
0.63C
0.92C
0.33C
0.60C
0.8C
0.33C
0.58C
0.71C
0.34C
0.56C
0.64C
-----------
C).
C error
N