The FIFO on the instrument is used as a circular buffer. Acquisition continues indefinitely
until you stop the operation. When the FIFO is full, the operation wraps to the beginning of
the FIFO; values are overwritten starting at the first location in the FIFO. It is up to your
application to retrieve the data from the FIFO; refer to your software documentation for more
information.
The conversion rate is determined by the frequency of the input sample clock; refer to
for more information about the input sample clock. The sample rate, which is the rate at which
a single entry in the channel list is sampled, is the same as the conversion rate due to the
simultaneous nature of the MEASURpoint instrument.
Figure 33
illustrates scanning a list of three enabled channels: channel 0, channel 1, and
channel 2. In this example, analog input data is acquired simultaneously on each clock pulse
of the input sample clock. Data is acquired continuously.
Input
Sample
Clock
Operation starts

Filtering

TEMPpoint, VOLTpoint, and MEASURpoint instruments use a Delta-Sigma analog-to-digital
converter (ADC) for each analog input channel to provide simultaneous sampling of all
inputs. The Delta-Sigma converter operates at 10 Hz effectively providing a filter that rejects
50 Hz and 60 Hz power line frequency components and that removes aliasing, a condition
where high frequency input components erroneously appear as lower frequencies after
sampling.
In addition to the filter provided in hardware, you can further reduce noise by selecting one of
the following filter options in software: Moving Average or Raw. Refer to your software
documentation for more information on selecting a filter type.

Data Format

TEMPpoint, VOLTpoint, and MEASURpoint instruments return data as 32-bit floating-point
values. How the data is returned depends on the channel type, as described in the following
subsections.
Chan 2
Chan 2
Chan 1
Chan 1
Chan 0
Chan 0
Data is acquired continuously
Figure 33: Continuous Scan Mode
Chan 2
Chan 2
Chan 1
Chan 1
Chan 0
Chan 0
Principles of Operation
page 95
97