OPERATION
MAINTENANCE
AND
INSTRUCTION MANUAL
AEU-7000E-70V & AEU-7000E
Implant / Endodontic Dental Systems
AEU-7000E-70V System

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AEU-7000E-70V and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Aseptico AEU-7000E-70V

  • Page 1 OPERATION MAINTENANCE INSTRUCTION MANUAL AEU-7000E-70V & AEU-7000E Implant / Endodontic Dental Systems AEU-7000E-70V System...
  • Page 2: Table Of Contents

    RX: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A DENTIST INDICATIONS FOR USE: The AEU-7000E-70V & AEU-7000E are drive systems for INFORMATION CONCERNING THE ACCURACY AND instruments and tools used in dentistry for implant/surgical PRECISION OF THIS PRODUCT MAY BE OBTAINED procedures and endodontic procedures.
  • Page 3: Introduction

    Your new Aseptico AEU-7000E-70V & AEU-7000E Systems are two of the finest dual- function implant/endodontic motor systems available to the dental profession. The systems combine a powerful brushless 40,000 RPM motor with a wide range of handpiece ratios and precision torque controls to make the perfect dental systems for Congratulations both implant and endodontic applications.
  • Page 4: Safety Precautions

    Important additional information and hints. SAFETY PRECAUTIONS: Aseptico accepts no liability for direct or consequential injury or damage resulting from improper use, arising in particular through the non-observance of the operating instructions, or improper preparation and maintenance of this product.
  • Page 5 Do not adjust the torque above the supplier-recommended rating or there is a risk of damage to the internal parts of your handpiece (which is not covered under warranty). Aseptico recommends and distributes AHP-85MB-series 20:1 handpieces, which are rated...
  • Page 6: Setting Up The Unit

    Fig.3 - Setup IRRIGATION BAG ELECTRONIC CONTROL HANGER ROD CONSOLE MULTI-FUNCTION FOOT CONTROL HANDPIECE & MOTOR CALIBRATION ADAPTER STORAGE MOTOR CRADLE & BRACKET DOOR RELEASE BUTTON DYNAMOMETER PORT IRRIGATION MOTOR RECEPTA- PUMP DOOR CORD CONNECTOR MOTOR CORD STRAIN RELIEF (AEU-7000E-70V System shown)
  • Page 7 b. Install Pump Tubing Assembly into pump 5. Attach the appropriate "E-Type" handpiece to door as shown in Figure 5. Install tubing the motor as shown in Figure 4. connector into the slot located on the Fig. 4 - Handpiece/Motor Connection back end of pump door.
  • Page 8: Control Panel Functions

    CONTROL PANEL FUNCTIONS: 1. Main Power Switch: press the foot pedal to wake up the System and return it to the last state used. Located on back of console (see Figure 2). Controls main power On / Off to the console. 3 Mode Select Button: The System will initialize with Implant Selects Implant or Endodontic...
  • Page 9: Speed Ranges

    5 Handpiece Ratio Selector: “20:1”, to “20.7:1”), indicating the precise ratio Allows user to select ratio of hand- measured. This also serves as an indicator piece. Ensures accurate display of that the handpiece has been calibrated. speed and torque levels. 6 Speed: a.
  • Page 10 CONTROL PANEL FUNCTIONS - Cont’d: 9 Torque Modes (A-S / MAX): No incremental adjustments are allowed Allows the user to select from when in “MAX” mode. one of three torque control NOTE: MAX Mode is only available with modes: Auto-Stop (Implant 1:1 and reduction handpieces.
  • Page 11 12 INSERT: the following procedures: Allows user to enter a blank Preset 1 - Site Preparation space into characters when Preset 2 - Pilot Drill editing the Preset button Preset 3 - Finish Drill / Reamer settings on the display. Preset 4 - Tap Forward 13 Display Edit &...
  • Page 12: Operation

    OPERATION: pressing the Standby button will return the GETTING STARTED: After the unit has been System to the last state used. NOTE: The set up and the user has become familiar with unit is in Sleep Mode when the Preset the System’s control panel functions, there LEDs blink consecutively.
  • Page 13 level reaches 75% and 100% of the Auto- handpiece and irrigation pump. Releasing Stop limit. Auto-Stop is the suggested the footswitch will stop the motor/handpiece mode when tapping and threading and pump. implants. 9. Calibration of Handpiece - Because b. MAX Torque Mode - Depressing the variations in handpiece efficiency can MAX (“A-S/ MAX”) button until the amber cause inaccuracies in torque, it is essential...
  • Page 14 OPERATION - Cont’d: Part-2 Calibration Procedure: Fig.8 - Adapter Installation If a reduction handpiece passes the “Free Run” calibration test above, the System automatically advances to the Part-2 “Dyno” calibration procedure. The following message will be displayed: Put Handpiece Into Dyno CALIBRATION ADAPTER Press 1>...
  • Page 15: Presets - Implant Mode

    OPERATION IMPLANT Presets Step 2: Press and hold any of the Preset buttons 1 through 6, to save the new, modified settings into that particular button. Step 3: The six preset memory buttons are A display prompt then asks the user: preprogrammed at the factory with the Preset - (X) default Implant Presets shown in Chart 1,...
  • Page 16 OPERATION IMPLANT Presets - Cont’d: Note: To enter a blank space Important: When the factory default into the text line, place the cursor settings are restored or recalled, or when under the character and press the unit has been reprogrammed with new the INSERT button.
  • Page 17: Presets - Endodontic Mode

    OPERATION ENDODONTIC Presets 13.Editing Endodontic Presets: All six preset memory buttons can be modified by the user with new files and operating parameters at any time. These new settings will overwrite the existing The six preset memory buttons are settings, including factory defaults. In preprogrammed at the factory with the addition, the new Preset Files’...
  • Page 18 OPERATION ENDODONTIC Presets, Cont’d: Step 4: Example: Press SELECT button to save “FILE NAME YYZ” settings or CANCEL out of the Press SELECT To Save menu. Step 5: Repeat Steps 8 & 9 for all remaining text characters that require edits. A new Preset entitled “User Defined Preset”...
  • Page 19 The scrolling “Help” message IMPORTANT: Whenever the unit’s factory instructs the user to press the default settings are recalled, or when the Up/Down arrow buttons to view unit has been reprogrammed with new two different menu choices: software, the default File Series will be restored to all the Preset buttons.
  • Page 20 OPERATION Chart 2 ENDODONTIC Default File Presets IMPORTANT Preset Button 1: PathFile The console will drive DENTSPLY Tulsa Dental Specialties PathFile All Files the files as close to the ® File #1: PathFile (All Files) library requested speed and torque as the Preset Button 2: ProTaper Next handpiece parameters DENTSPLY Tulsa Dental Specialties ProTaper Next™...
  • Page 22: System Setup

    SYSTEM SETUP warns the user with an audible signal The SETUP Program allows the when a specified Torque Limit is reached. user to select/configure setup Warning signals are provided differently, options via display prompts. The depending upon which operating mode is option selections and corresponding activated.
  • Page 23 b. To enable a 30-minute delay, press 4. Auto Stop Mode (for Endo Mode Only): Preset Button #2. When the Auto Stop Reverse (ASR) feature is enabled, rotation of the c. To disable Sleep Mode (Display stays handpiece will automatically stop and on), press Preset Button #3.
  • Page 24: Foot Switch Description & Operation

    Fig.11 - Variable-Speed Foot Control comes as standard equipment on the HANDLE (Removable) AEU-7000E-70V System and as an option on the AEU-7000E System. The AE-70V2 can control motor speed, direction, torque, TORQUE MODE and turn the pump On/Off. It can select...
  • Page 25 Handle Installation/Removal: files, press and then quickly release the pad repeatedly. 7. The Foot Switch Handle may be installed to Note: The following message will appear allow the user to reposition or move the briefly whenever any Endo Preset is Foot Control more easily.
  • Page 26: Reprogramming The Unit

    Programming Successful. devices. These cards, available from Eject Card. Aseptico, enable a user to update software or replace existing software that might • Press the card inward slightly, then have been accidentally erased or release it to eject it.
  • Page 27: Sterilization

    Wrapped Sterilization – Place in an The entire motor & appropriately sized sterilization pouch and cord assembly is seal it. autoclavable. NOTE: Call Aseptico Inc. at 1-800-426-5913 A. Standard autoclaving (Gravity displacement method) for any questions or clarifications on this Time: 15 min sterilization procedure.
  • Page 28: Maintenance & Cleaning

    MAINTENANCE & CLEANING: SPECIFICATIONS: ANDPIECES - Thorough cleaning and lubrication Console Dimensions: 9.98”W x 9.42”L x 5.10”H (25.3 cm x 23.9 cm x 12.9 cm) of handpieces after each use and before sterilization is very important to ensure proper operation and Console Weight: 7.3 lbs (3.3kg) service life of the handpiece.
  • Page 29: Troubleshooting

    Then, grasp the connector body near the red dot and pull the connector straight out of the motor receptacle. CHANGING THE FUSE: W RNING NOTE: The AEU-7000E-70V & AEU-7000E feature auto-sensing, global voltage compatibility. The fuse indicated is correct for Turn the power off and unplug the unit 100V-240V 50/60 Hz line voltage.
  • Page 30: Symbol Definitions

    SYMBOL DEFINITIONS: Authorized European Attention, consult Representative accompanying documents Type BF Equipment Standby Switch Dangerous Voltage Footswitch Alternating current Do Not Throw Into Trash Protection Against Dripping Water Manufacturer Protective Earth (Ground) Fuse Rating Motor Direction Preset Step Through Pump On/Off Torque Step Through Temperature Limitation Atmospheric Pressure Limitation...
  • Page 31 NOTES:...
  • Page 32: Warranty

    Aseptico's sole obligation under product warranty is (at its sole option and discretion) to repair or replace any defective component or product in part or whole. Aseptico shall be the sole arbiter of such action.