Fitting The Spindle; Fitting A Sample Container Support - Elcometer 2250 Operating Instructions Manual

Krebs viscometer
Table of Contents

Advertisement

3.4 Fitting the spindle

Note: This operation must be done very
carefully to ensure that the spindle is not
bent and the motor drive coupling or internal
components are not damaged.
1. Check that the connecting surfaces of
the spindle and the motor drive
coupling are clean.
2. Insert the spindle as far as it will go up
into the motor drive coupling.
3. Using the hexagonal wrench supplied,
rotate the retaining screw until it just
makes contact with the spindle.
4. Pull down on the spindle until its
movement is stopped by the retaining
screw.
5. Tighten the retaining screw.
3.5 Fitting a sample container
support
During a test, the sample container (glass
beaker or can) is held in position at the
correct height by the sample container
support. Two supports are supplied with the
instrument - a large one and a smaller one.
Which support you use depends upon the
size of the sample container.
The sample container support is held in
position on the base by the locating plug.
To fit the sample container support:
When using the glass beaker supplied
with the instrument, or a 1 pint (USA)
can
1. Place the locating plug into the recess
on the base of the instrument.
2. Place the large sample container
support onto the locating plug.
When using a ½ pint (USA) can
1. Place the locating plug into the recess on the base of the instrument.
2. Place the large sample container support onto the locating plug.
3. Place the smaller sample container support into the recess on the top of the large sample
container support.
8
R
Drive
coupling
Spindle
Smaller
support
Large
support
Locating
plug