The instrument displays "Going DOWN" and the head moves down until the spindle reaches
its lowest position.
When the spindle reaches its lowest position, the sample wait timer starts to count down.
When the sample wait timer reaches zero, the spindle starts rotating and the sample
measurement timer starts to count down. While the spindle rotates, the instrument displays the
viscosity reading:
When the sample measurement timer reaches zero, the spindle stops rotating, the head
moves up and the instrument displays the final viscosity reading and the option to print the
results or quit:
9. If necessary, press [ENTER] to print the results. If
your Viscometer is connected to a suitable printer,
this provides a quick and easy way to save and
document individual test results. See "Spare parts
and accessories" on page 17.
The results from the test are output via the RS232
connection at the rear of the Viscometer head.
10. When you have finished taking the measurement,
press [STOP].
11. After use, clean the instrument and spindle - see
"Cleaning" on page 16.
automatic process
SWT: 05s. SMT: 10s
Ku: 000.0
Going DOWN
automatic process
SWT: 05s. SMT: 10s
Ku: 121
<STOP> quit
automatic process
SWT: 05s. SMT: 10s
Ku: 121
<ENT> print <STP> quit