Honeywell EK280 Operating Instructions Manual page 70

Volume conversion device
Hide thumbs Also See for EK280:
Table of Contents

Advertisement

70
6.3.6.1
Calibration lock
The calibration lock is used to protect parameters subject to calibration
regulations. This includes all values which influence volume readings or the
volume conversion.
The calibration lock is designed as a button which is positioned inside the
EK280 housing underneath the circuit board cover. It can be protected with
an adhesive label (see chapter 5.3.1.1).
The calibration lock is opened by pressing the button ("P" symbol flashes in
the display) and is closed again by pressing the same button ("P" symbol
fades). It can also be closed by deleting the value "St.PL" (see chapter
6.4.6) via the keyboard or interface. Using the configuration software
"enSuite", a time can also be set (in minutes) after which the calibration
lock will automatically close.
The level of protection can be changed for all parameters upon request,
particularly for applications which are not subject to German calibration
guidelines or the MID.
For example, parameters which are protected by the calibration lock as
standard, can also be protected with a administrator's lock or a certification
data log.
6.3.6.2
Certification data log
The certification data log is activated as standard but can also be disabled.
The corresponding parameters will then be protected by the calibration
lock.
Using the "certification data log" as per PTBA 50.7, several parameters
subject to calibration regulations can also be adjusted when the calibration
lock is closed. This is subject to the following conditions:
The administrator's lock (see below) must be open.
There are at least three free entries in the certification data log.
The following parameters are set as default:
"cp.I1" cp value for input 1
"MPer" measurement period
"Md.K" K-factor mode
"Clr.A" delete measurement archive
"Md.I1" mode for input 1
"qMax" maximum flow (plausibility check in encoder mode)
For every change of a parameter listed under "CDL" when the calibration
lock is closed, one row of data will be entered for the value before and after
the change.