56
Serv.
Measured values
and pMax
Serv.
Measured values
TMin and TMax
–
Press the ENTER button to confirm the input.
Use the arrow keys or to go to the digits and change these with
–
the arrow keys or .
–
Press the ENTER button to confirm the set value. The input value
can be deleted by pressing the ESC button.
5.3.1.12 Adjusting replacement values for gas pressure and gas
temperature
If the gas pressure or gas temperature cannot be measured due to a fault,
an adjustable replacement value (fixed value) will be used to calculate the
volume at base conditions. The volume calculated in this way will be
counted in the separate disturbance quantity meter "VbD".
The replacement values can be found under the following paths:
Serv.
Measured values
Serv.
Measured values
Changing the values generally takes place precisely as described in
chapter 5.3.1.12 for the alarm limit values.
5.3.1.13 Configuration of the outputs
The cp values (pulse constants) for the outputs can also be
configured using the enSuite software.
The following functions are preset ex-factory for the outputs:
Outputs
Function
Output 1
Pulse output for the total volume at base conditions "VbT"
(cp.O1 = 0.1)
Output 2
Pulse output for the total volume at measurement conditions
"VmT" (cp.O2 = 0.1)
Output 3
Status output for warnings (see chapter 8.2)
Output 4
Status output for alarms (see chapter 8.2)
The cp value of an output indicates how many pulses are emitted per cubic
meter. A cp value of 0.1 / m3 (0.1 pulses per m3) means, for example, that
one pulse is emitted per 10 m3.
Assembly, Connection and Putting into Operation
Pressure
Temperature
Pressure
Temperature
Parameter settings
Parameter settings
Parameter settings
Parameter settings
pMin
p.F
T.F