4 Installation
4.4 Initial start-up – procedure for proper assembly
34
WARNING
Use of the laser unit when the aiming beam is not functioning properly
may cause injuries to operating personnel, assistants or patients. If
you cannot see the red aiming beam after switching the laser on or
during treatment refer to chapter "Troubleshooting of simple defects 
[→ 68]".
Only use the MultiTips for treatments appropriate for them. EasyTips
and MultiTips have completely different optical characteristics.
Remove the protective cap of the handpiece only for treatment.
Never touch the proximal end of the cap and protect it against
damage and dirt.
Never use the laser without optical fiber or MultiTip light guide, check
for correct fixation.
After treatment
Disconnect the MultiTip from the handpiece by removing it carefully
from the optical socket of the handpiece.
CAUTION
As soon as you disassemble the MultiTip after treatment make sure to
protect the optical fiber socket with the protective cap provided for this
purpose. Make sure that no dust or dirt can enter the optical system.
Otherwise the unit may be permanently damaged.
For cleaning, disinfecting and sterilizing the MultiTip please refer to
chapter "Cleaning, disinfection and sterilization  [→ 63]".
4.4.3
Install wireless foot control – optional
The SiroLaser Blue can be operated using the finger switch (which is
integrated in the handpiece) as well as by using the optional wireless
foot control.
NOTE
The foot switch has an IPX5 degree of protection. Therefore this foot
switch is not suitable for use in hospital operating rooms.
For technical data of the wireless foot control, see chapter Technical
Data, "Wireless foot control".
The wireless foot control must be assigned to the SiroLaser Blue via a
registration. This prevents malfunctions caused by neighboring wireless
controls.
ü The SiroLaser Blue control unit and the wireless foot control are
ready for operation.
1. Choose "Settings" from the main home screen.
2. Choose "Activation device".
3. Choose "Wireless registration".
4. Press the foot pedal for 3 seconds and follow the instructions on the
screen.