SE-3&SE-300 Series Electrocardiograph User Manual
External Input
External Output
Language
System Password
Load Factory
Settings
The external input socket is equipped in the electrocardiograph, through
which the electrocardiograph can receive signals from the external
equipment.
Choose from: On or Off, the default value is Off
Select On, the electrocardiograph will display the signals which it
receives from external input port.
The external output socket is equipped in the electrocardiograph, through
which the electrocardiograph can send rhythm lead signals to the external
equipment.
Choose from: Off, Standard or Triggered, the default value is Off
Select Standard, the electrocardiograph sends ECG signals of rhythm
lead 1.
Select Triggered, the electrocardiograph sends pulses with the height of
5V and the width of 45ms, based on the data of rhythm lead 1.
Select the language displayed on the main screen and in the ECG reports.
Type a password that allows you to access the System Setup window.
Press to restore the factory settings.